Product Specifications Measurement Specifications
TM80 Telemetry Monitor Operator’s Manual A - 17
A.8.3.4 Fitting Curve for Nellcor SpO
2
Sensors
â–  SpO
2
Accuracy for Nellcor Sensors vs. Co-Oximeters (Arms)
â–  Modified Bland-Altman for SpO
2
-DS-100A Sensor: (SpO
2
- SaO
2
) vs. SaO
2
SpO
2
Range 100% to 70% 100% to 90% 89% to 80% 79% to 70%
DS-100A 1.64% 1.16% 1.67% 2.25%
D-YS, OXI-P/I 2.41% 1.38% 2.50% 3.60%