VNA GUI reference
R&S
®
ZNL/ZNLE
485User Manual 1178.5966.02 ─ 20
Same Connector All Ports / Same Gender All Ports
Assigns the same connector type or gender to all selected physical ports. For some
multi-port calibration types, the port connector types must be equal, e.g. because they
require a Through standard with known characteristics.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:CONNection:PORTs
[SENSe<Ch>:]CORRection:COLLect:CONNection:GENDers
Import Cal Kit...
Opens the "Import Calibration Kit" dialog that allows you to import a cal kit file. For
background information, see Chapter 9.5.2.4, "Cal kit Files", on page 260.
Remote command:
MMEMory:LOAD:CKIT
Back
Go back to Step 1: Ports and Type.
Start
Start Step 3: Calibration.
Step 3: Calibration
Allows you to acquire error correction data for every required port (pair) and calibration
standard, where "required" depends on the selected ports and calibration type. On
"Apply", the R&S ZNL/ZNLE calculates the system error correction (error terms) from
the measurement data of the standards and applies the result to the active channel.
Cal softtool