VNA GUI reference
R&S
®
ZNL/ZNLE
521User Manual 1178.5966.02 ─ 20
10.12.4.1 Controls on the Use Cal tab
Figure 10-22: Use Cal tab: R&S
ZNL (left) vs. R&S
ZNLE (right)
User Cal Active
Activates or deactivates the system error correction in the active channel. "User Cal
Active" is available only if a valid system error correction is available for the active
channel; see "Channel State" in Chapter 10.12.4.3, "Calibration Manager dialog",
on page 523.
Note: A label "Cal Off" appears behind the trace list if the system error correction is
switched off; see also Chapter 9.5.4, "Calibration state labels", on page 262. The cali-
bration status of each channel and trace appears in the setup information ("Setup" >
"Info..." > "Setup").
Remote command:
[SENSe<Ch>:]CORRection[:STATe]
Scalar Power Cal – All Power Cals On / All Power Cals Off
Activates or deactivates all scalar power calibrations in the active channel. "All Power
Cals On" is only available if a valid power calibration is available for the active channel,
but not active; see "Channel State" in Chapter 10.12.4.3, "Calibration Manager dialog",
on page 523.
Note: A label "PCal Off" appears behind the trace list of a wave quantity or a ratio if the
power calibration is switched off; see also Chapter 9.5.6.3, "Power calibration labels",
on page 275.
Remote command:
[SENSe<Ch>:]CORRection:PCAL
Active Power Cals...
Opens the Active Power Cals dialog
Cal softtool