VNA GUI reference
R&S
®
ZNL/ZNLE
522User Manual 1178.5966.02 ─ 20
Manage Cals – Recall Last Cal Set
Loads and activates the channel setup for which the last calibration was performed. If
the last calibrated setup is already active, nothing is changed. The calibrated setups
are automatically stored in the
C:\Users\Public\Documents\Rohde-Schwarz\ZNL\Calibration\
RecallSets directory. A message box pops up if the directory is empty, e.g. because
no calibration was performed yet.
Remote command:
n/a
Manage Cals – Cal Manager...
Opens the Calibration Manager dialog
Manage Cals – Remove all Cal Data
This button cleans the active channel's existing calibration data and also removes links
to the calibration pool.
Remote command:
[SENSe<Ch>:]CORRection:DELete
10.12.4.2 Active Power Cals dialog
The "Active Power Cals" dialog shows the power calibrations that are compatible with
the active channel configuration. It allows you to enable or disable the available power
calibrations.
Access: [Cal] > "Use Cal" > "Active Power Cals..."
All possible power calibrations can be performed from the Power Cal wizard.
Port Overview
Shows all source ports together with their possible power calibrations. Unavailable cali-
brations (i.e. calibrations that have not been performed yet) are grayed out.
Cal softtool