EasyManua.ls Logo

R&S ZNL Series - Page 940

R&S ZNL Series
1135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNL/ZNLE
940User Manual 1178.5966.02 ─ 20
Parameters:
<Boolean> Enables (ON) or disables (OFF) the receiver power calibration
for the received waves b
<PhyPt>
.
*RST: OFF
Example:
*RST; CORR:POW:ACQ BWAVe,1,PORT,2
Perform a receiver power calibration of the wave b1 using port 2
as a source port, assuming that the source power a
2
is correct.
A through connection from port 2 to port 1 is needed.
CORR:POW?
Check whether the calibration is applied (the response is 1).
CALC:PAR:MEAS 'TRC1', 'B1D2'
Select b
1
as a measured quantity for the default trace.
CALC:MARK ON; MARK:Y?; :SOUR:POW?
Create marker no. <Mk> in the center of the sweep range and
query the measurement value. The calibrated power of the
received wave b
1
is approx. equal to the default source power
value.
See also example for [SENSe<Ch>:]CORRection:
POWer<PhyPt>:ACQuire.
Manual operation:
See "Port Overview" on page 522
[SENSe<Chn>:]CORRection:PSTate?
Gets the power calibration label of the active trace in channel <Chn>.
See Chapter 9.5.6.3, "Power calibration labels", on page 275 for the possible return
values.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Usage: Query only
Manual operation: See "Apply" on page 490
[SENSe<Ch>:]CORRection[:STATe] <Boolean>
Enables or disables the system error correction for channel <Ch>.
Suffix:
<Ch>
.
Calibrated channel number
Parameters:
<Boolean> Enables (ON) or disables (OFF) the correction.
*RST: ON
Example:
*RST; :CORR?
Reset the instrument and query whether channel 1 is system
error corrected. The response is 1.
Manual operation: See "User Cal Active" on page 521
VNA command reference

Table of Contents

Related product manuals