CALCulate:CALValidate:RESPonse:IMAGinary:LOWer <Port>, <Value>
CALCulate:CALValidate:RESPonse:IMAGinary:UPPer <Port>, <Value>
CALCulate:CALValidate:RESPonse:REAL:LOWer <Port>, <Value>
CALCulate:CALValidate:RESPonse:REAL:UPPer <Port>, <Value>
Defines the upper and lower cal validation limits for the REAL trace format (see
CALCulate:CALValidate:FORMat).
Use CALCulate:CALValidate:CHARacteriza to select the reference characteri-
zation.
Parameters:
<Port>
<Value> *RST: -200 (LOWer) | +200 (UPPer)
Default unit: mU
Manual operation: See "Validation Limits" on page 495
CALCulate:CALValidate:RESPonse:MAGNitude:LOWer <Port>, <Value>
CALCulate:CALValidate:RESPonse:MAGNitude:UPPer <Port>, <Value>
Defines the magnitude cal validation limits for the MLINear trace format (see
CALCulate:CALValidate:FORMat).
Use CALCulate:CALValidate:CHARacteriza to select the reference characteri-
zation.
Parameters:
<Port> Physical analyzer port no.
<Value> *RST: -0.2 (LOWer) | +0.2 (UPPer)
Default unit: dB
Manual operation: See "Validation Limits" on page 495
CALCulate:CALValidate:RESPonse:PHASe:LOWer <Port>, <Value>
CALCulate:CALValidate:RESPonse:PHASe:UPPer <Port>, <Value>
Defines the phase cal validation limits for the MLINear trace format (see CALCulate:
CALValidate:FORMat).
Use CALCulate:CALValidate:CHARacteriza to select the reference characteri-
zation.
Parameters:
<Port>
<Value> *RST: -2 (LOWer) | +2 (UPPer)
Default unit: degrees
Manual operation: See "Validation Limits" on page 495