VNA GUI reference
R&S
®
ZNL/ZNLE
495User Manual 1178.5966.02 ─ 20
Use "Detect Port Assignment" to detect the current port assignment (after reconnecting
the calibration unit).
Measurement
Selects the measurement to be performed in advanced mode and runs it
Characterization
Selects the calibration unit characterization to be used for the validation.
By default the factory characterization is used.
Remote command:
CALCulate:CALValidate:CHARacteriza
Calibration Standard
Selects the calibration standard to be measured in advanced mode.
Validation Limits
The "Limits..." button opens the "Cal Validation Limits" dialog that allows you to specify
the maximum tolerable deviations from the corresponding S-parameters of the
selected reference characterization.
Figure 10-19: Validation limits for Mag/Phase format
Depending on the selected
Format, different limits can be set.
"Mag/Phase": lower and upper limits for magnitude and phase; defaults are
±0.2 dB for magnitude and ±2° for phase deviations.
"Imag/Real": lower and upper limits for real and imaginary part; defaults are
±200 mU for both of them.
Remote command:
CALCulate:CALValidate:RESPonse:MAGNitude:LOWer
CALCulate:CALValidate:RESPonse:MAGNitude:UPPer
CALCulate:CALValidate:RESPonse:PHASe:LOWer
CALCulate:CALValidate:RESPonse:PHASe:UPPer
CALCulate:CALValidate:RESPonse:REAL:LOWer
CALCulate:CALValidate:RESPonse:REAL:UPPer
CALCulate:CALValidate:RESPonse:IMAGinary:LOWer
CALCulate:CALValidate:RESPonse:IMAGinary:UPPer
CALCulate:CALValidate:RESPonse:PRESet
Preferences
Allows you to configure the behavior of the validation diagrams and traces.
Cal softtool