Command reference
R&S
®
ZNL/ZNLE
840User Manual 1178.5966.02 ─ 20
Parameters:
<Boolean> ON - the analyzer measures continuously in all channel, repeat-
ing the current sweep. The query returns ON (1) if at least one
channel is measured continuously.
OFF - the measurement is stopped after the number of sweeps
defined via [SENSe<Ch>:]SWEep:COUNt. INITiate<Ch>[:
IMMediate][:DUMMy] initiates a new measurement cycle.
*RST: ON
Example: See INITiate<Ch>:CONTinuous
Manual operation: See "All Channels Continuous/All Channels on Hold"
on page 344
INITiate[:IMMediate]:ALL
Starts a new single sweep sequence in all channels. This command is available in sin-
gle sweep mode only (INITiate:CONTinuous:ALL OFF). The data of the last sweep
(or of previous sweeps, see Retrieving the results of previous sweeps) can be read
using CALCulate<Chn>:DATA:NSWeep:FIRSt? SDATa, <count>.
This command is not supported in the "ZVR" or "ZVABT" compatibility modes.
Note: In contrast to all other commands of the VNA mode, the
INITiate<Ch>[:IMMediate]... commands have been implemented for overlap-
ped execution.
Example:
See INITiate<Ch>:CONTinuous
Usage: Event
Manual operation: See "Restart Sweep" on page 344
INITiate<Ch>[:IMMediate][:DUMMy]
Starts a new single sweep sequence. This command is available in single sweep mode
only (INITiate<Ch>:CONTinuous OFF). The data of the last sweep (or previous
sweeps, see "Retrieving the results of previous sweeps" on page 1073) can be read
using CALCulate<Chn>:DATA:NSWeep:FIRSt? SDATa, <count>.
Note: In contrast to all other commands of the VNA mode, the
INITiate<Ch>[:IMMediate]... commands have been implemented for overlap-
ped execution.
Suffix:
<Ch>
.
Channel number. If the channel number does not exist the ana-
lyzer returns an error message. If the "ZVR" or "ZVABT" compat-
ibility mode is active and INITiate<Ch>[:IMMediate]:
SCOPe ALL is selected, this suffix is ignored.
Example: See INITiate<Ch>:CONTinuous
Usage: Event