Command reference
R&S
®
ZNL/ZNLE
862User Manual 1178.5966.02 ─ 20
EMBedding<PhyPt>:TNDefinition to select the adequate circuit model before
executing this command.
●
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
PARameters:DATA to load circuit data from the remote client.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<TouchstoneFile> String parameter to specify the name and directory of the loaded
Touchstone file.
If no path is specified the analyzer searches the current direc-
tory, which can be set and queried using MMEMory:
CDIRectory?
<Interchange> FPORts | IPORts | SGATes
FPORts (or omitted)
Standard port sequence (network port 1 towards VNA, network
port 2 towards DUT)
IPORts | SGATes
Inverted port sequence (network port 2 towards VNA, network
port 1 towards DUT)
Example:
CALC:TRAN:VNET:SEND:EMB:TND FIMP
Select the "Serial .s2p" data circuit model.
MMEM:LOAD:VNET:SEND:EMB2 'C:\Users\Public
\Documents\Rohde-Schwarz\ZNL\VNET\Test.s2p'
Load a Touchstone file and assign it to the physical port no. 2.
Manual operation: See "File Name 1 / Swap Gates" on page 536
MMEMory:STORe:CABLe <OutputDir>
Saves all predefined and user-defined Distance to Fault (DtF) cable types to the speci-
fied output directory.
Each DtF cable type is stored in a separate ASCII file <cable name>.rsc. The com-
mand silently overwrites files of the same name previously existing in the output direc-
tory.
Setting parameters:
<OutputDir> Output Directory, must be created before the command is exe-
cuted
Example: see MMEMory:LOAD:CABLe on page 845
Usage: Setting only
Options: R&S ZNL-K3
Manual operation: See "Add / Delete" on page 374
VNA command reference