Command reference
R&S
®
ZNL/ZNLE
912User Manual 1178.5966.02 ─ 20
[SENSe<Ch>:]CORRection:COLLect:CKIT:PORT<PhyPt>? [<Detail>]
Queries the cal kit data assigned to the given port via
[SENSe<Ch>:]CORRection:COLLect:CKIT:LOAD
or [SENSe<Ch>:]CORRection:COLLect:CKIT:INSTall
Suffix:
<Ch>
.
Channel number
<PhyPt> Port number
Query parameters:
<Detail> CONNector | LABel | NAME | GENDer
Queried property
Return values:
<Result>
Usage: Query only
[SENSe<Ch>:]CORRection:COLLect:CONNection<PhyPt> <ConnectorType>
Selects a connector type at a specified port <PhyPt> and its gender.
Tip: Use [SENSe<Ch>:]CORRection:COLLect:SCONnection<PhyPt> to select
an arbitrary connector type using a string variable.
Suffix:
<Ch>
.
Channel number of the calibrated channel
<PhyPt> Port numbers of the analyzer
Note: If the analyzer is set to use the same connectors at all
ports ([SENSe<Ch>:]CORRection:COLLect:CONNection:
PORTs ALL), then a change of a connector type is valid for all
ports. The gender of the connectors can still be different.
VNA command reference

Table of Contents

Other manuals for R&S ZNL Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNL Series and is the answer not in the manual?

R&S ZNL Series Specifications

General IconGeneral
BrandR&S
ModelZNL Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals