Command reference
R&S
®
ZNL/ZNLE
975User Manual 1178.5966.02 ─ 20
<Grp> Port group number.
This suffix is ignored: only <Grp>=1 is supported.
Setting parameters:
<arg0> ALL
If ALL is specified, all port groups are dissolved and a default
port group 1, consisting of all available ports, is restored; the
<Grp> suffix is ignored.
If ALL is omitted, only the specified port group is dissolved,
except the deleted port group was the only one, in which case
again default port group 1 is restored. In case an undefined port
group number is used, the analyzer generates an error mes-
sage.
Example: See SOURce<Ch>:LPORt<LogPt> on page 997
Usage: Setting only
SOURce<Ch>:GROup<Grp>:PORTs
<port0>[, <port1>[, ...]]
Defines the set of active logical ports (the "master group") as an arbitrary selection of
logical ports. The ports do not have to be numbered consecutively (as for port groups
defined via SOURce<Ch>:GROup<Grp>).
See SOURce<Ch>:LPORt<LogPt> on how to define logical ports.
Suffix:
<Ch>
.
Channel number.
<Grp> Port group number, must be 1 (or omitted).
Parameters:
<port0>, <port1>, ... Logical port numbers. Must be omitted if the command is used
as a query.
SOURce:POWer:CORRection:COLLect...
The SOURce:POWer:CORRection:COLLect... commands allow you to set up a
scalar source power calibration.
SOURce:POWer:CORRection:COLLect:AVERage[:COUNt] <NoReadings>
Sets a limit for the number of calibration sweeps in the source power calibration. The
command is valid for all channels and ports.
Equivalent command: SOURce:POWer:CORRection:NREadings.
Parameters:
<NoReadings> Number of readings
Range: 1 to 100
*RST: 2
VNA command reference