Command reference
R&S
®
ZNL/ZNLE
976User Manual 1178.5966.02 ─ 20
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire] (for
equivalent command SOURce:POWer:CORRection:
NREadings)
Manual operation: See "Flatness Cal – Max Iterations" on page 510
SOURce:POWer:CORRection:COLLect:AVERage:NTOLerance <Tolerance>
Specifies the maximum deviation of the measured power from the target power of the
calibration. The command is valid for all channels and calibrated ports.
Parameters:
<Tolerance> Tolerance value
Range: 0.001 dB to 1000 dB
*RST: 1 dB
Default unit: dB
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation: See "Flatness Cal – Tolerance" on page 511
SOURce:POWer:CORRection:COLLect:CFACtor <Convergence>
Specifies the convergence factor for a source power calibration.
Parameters:
<Convergence> Convergence factor
Range: 0 to 2
Increment: 0.1
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation: See "Flatness Cal – Convergence" on page 511
SOURce:POWer:CORRection:COLLect:FLATness <Boolean>
Enables or disables the source power calibration. The command is valid for all chan-
nels and ports.
It is not possible to disable flatness calibration and the reference receiver calibration
simultaneously (see SOURce:POWer:CORRection:COLLect:RRECeiver
on page 977).
Parameters:
<arg0> <Boolean>
Disables (OFF|0) or enables (ON|1) the flatness calibration. With
disabled flatness calibration, only one calibration sweep is per-
formed in order to calibrate the reference receiver; the previous
source calibration data is not overwritten.
*RST: ON|1
VNA command reference