Command reference
R&S
®
ZNL/ZNLE
985User Manual 1178.5966.02 ─ 20
SOURce<Ch>:POWer:CORRection[:ACQuire]..................................................................985
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]:VERification:RESult?...................... 986
SOURce<Ch>:POWer:CORRection:DATA........................................................................987
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?............................................ 988
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?.....................989
SOURce:POWer:CORRection:NREadings........................................................................989
SOURce<Ch>:POWer:CORRection:OSOurces[:STATe]..................................................... 989
SOURce:POWer:CORRection:PMETer:ID........................................................................ 990
SOURce:POWer:CORRection:PPOWer........................................................................... 990
SOURce:POWer:CORRection:PSELect........................................................................... 991
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe.......................................................... 991
SOURce<Ch>:POWer:CORRection[:ACQuire] <SourceType>[, <SourcePort>[,
<CalOnlyPortFreq>]]
Selects the source for the source power calibration, starts and applies the source
power calibration.
Note: The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Suffix:
<Ch>
.
Calibrated channel number
Setting parameters:
<SourceType> PORT | A1 | A2
PORT – analyzer port is the source of the calibrated wave.
The port number is specified by means of the <SourcePort>
parameter.
A1 | A2 – direct analyzer port selection. The parameters A1 and
A2 are ZVR-compatible. No additional port number parameter
needs to be specified.
<SourcePort> Analyzer port number, if PORT is the first parameter.
<CalOnlyPortFreq> Relevant for frequency conversion measurements only.
OFF | 0
The reference receiver at port <CalPort> is calibrated for all fre-
quencies that are relevant for any of the ports.
ON | 1
Only those frequencies are calibrated that are relevant for this
port. This results in shorter calibration times.
*RST: 0
VNA command reference