Device-specific, command or query
See SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire].
SOURce<Ch>:POWer<Pt>:CORRection:OSOurces[:STATe] <Boolean>
Switches off all other sources during the calibration sweep for channel <Ch>.
Calibrated channel number
Calibrated port number. This parameter is ignored; the setting is valid for all
sources.
ON – Other sources not necessarily switched off
OFF – Other sources switched off during the calibration sweep
Device-specific, command or query
See SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire].
SOURce<Ch>:POWer<Pt>:CORRection:PMETer:ID <pmeter_no>
Selects an external power meter for the source power calibration.
The command cannot be used unless a power meter is connected via GPIB bus, USB or LAN
interface and configured in the External Power Meters dialog.
Calibrated channel number
Calibrated port number. This parameter is ignored; the power meter selection is
valid for all sources.
Number of external power meter. The parameters UP, DOWN, MIN, MAX are
not available for this command.
1 to number of external generators [–]
The power meter selection is not changed by a reset of the analyzer.
Device-specific, command or query.
See SOURce<Ch>:POWer<Pt>:CORRection:COLLect[:ACQuire] and
SOURce<Ch>:POWer<Pt>:CORRection:MIXer:RF[:ACQuire].
SOURce<Ch>:POWer<Pt>:CORRection:STATe <Boolean>
Enables or disables the source power calibration for channel <Ch> and for port number <Pt>. The setting
command is disabled unless a source power calibration for the analyzer port has been performed
(SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire]). The query always returns a result. To enable or
disable a source power calibration for an external generator use
SOURce<Ch>:POWer<Pt>:CORRection:GENerator[:STATe].