R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 1054
SYSTem:COMMunicate:RDEVice:AKAL:SANitize
Securely erases all user characterizations from the internal memory of the active calibration unit
(SYSTem:COMMunicate:RDEVice:AKAL:ADDRess).
SYST:COMM:RDEV:AKAL:ADDR 'ZN-Z51::900001'
Make the cal unit named 'ZN-Z51::900001' the active cal unit.
SYSTem:COMMunicate:RDEVice:AKAL:SANitize
Perform the sanitization.
SYSTem:COMMunicate:RDEVice:EUNit:IDN?
Queries the device ID of a connected extension unit R&S ZVAX24.
Manufacturer, type of the extension unit, 10-digit order number, serial number, see
example below. If no extension unit is connected, an execution error is generated.
– (*RST has no effect on external devices)
Device-specific, query only
SYST:COMM:RDEV:EUN:IDN?
Query the device ID. A possible device response is
Rohde&Schwarz,ZVAX24,1311250902100001, where 1311.2509.02 is the order
number of the extension unit R&S ZVAX24, 100001 is the serial number of the
connected unit.
SYST:COMM:RDEV:EUN:OPT?
Query the installed hardware options of the extension unit. A possible device
response is ZVAX-B211, ZVAX-B252, ZVAX-B273.
SYSTem:COMMunicate:RDEVice:EUNit:OPT?
Queries the hardware options of a connected extension unit R&S ZVAX24.
Comma-separated list of options, see example.
– (*RST has no effect on external devices)
Device-specific, query only
See SYSTem:COMMunicate:RDEVice:EUNit:IDN?