R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 1057
Number of the configured power meter. Power meters must be numbered in
ascending order, starting with 1. If a number is re-used, the previous power meter
configuration is overwritten.
Power meters can be assigned several times so that the number of configured
power meters is practically unlimited.
*RST has no effect on the configured power meters.
Device-specific, no query
See SYSTem:COMMunicate:RDEVice:PMETer<pmeter_no>:DEFine
SYSTem:COMMunicate:RDEVice:PMETer<pmeter_no>:CATalog?
Queries the numbers of all configured external power meters.
Number of the configured power meter. This suffix is ignored; the command
affects all power meters.
String parameter with a comma-separated list of power meter numbers.
*RST has no effect on the configured power meters.
Device-specific, query only
SYST:COMM:RDEV:PMET1:DEF 'USB Power Meter 1', 'NRP-Z55', 'usb', '100045'
SYST:COMM:RDEV:PMET2:DEF 'USB Power Meter 2', 'NRP-Z55', 'usb', '100046'
Configure two R&S NRP power meters as external power meter no. 1 and 2,
assigning names and serial numbers.
SYST:COMM:RDEV:PMET1:CAT?
Query the power meter numbers. The response is '1, 2'.
SYST:COMM:RDEV:PMET1:COUN?
Query the number of configured power meters. The response is 2.
SYSTem:COMMunicate:RDEVice:PMETer<pmeter_no>:CONFigure:AUTO[:STATe]
<Boolean>
Enables or disables Auto Config NRP-Zxx. If the function is enabled, the analyzer automatically configures
all power meters detected at any of the USB ports as Pmtr 1, Pmtr2, ...
Number of the configured power meter. This suffix is ignored; the command
affects all power meters.
Power meters can be assigned several times so that the number of configured
power meters is practically unlimited.
ON | OFF – Enable or disable Auto Config NRP-Zxx
*RST has no effect on the configured power meters.
Device-specific, command or query