R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 1061
*RST has no effect on the calibration algorithm.
Device-specific, command or query
SYST:CORR:FMP ON
Enable fast multiport correction.
SYSTem:DATA:SIZE ALL | AUTO
Defines the block size for the data transfer between the instrument hardware and the software. The block
size has an impact on the display of traces and on the readout of trace data using the commands of the
TRACe... or CALCulate<Chn>:DATA... subsystems.
Data transfer with full buffer size. The data for each sweep are transferred together
so that each displayed trace contains the data of exactly one sweep. The same is
true for queried trace data. The trace is displayed or updated at once, the sweep
progress is not visible on the screen.
To make sure that all queried trace points stem from the same sweep you can
also use single sweep mode (INITiate<Ch>:CONTinuous OFF) and the common
command *OPC?
Data transfer with an automatically determined block size. In general the data blocks
comprise only part of a trace, therefore the sweep progress is visible on the screen.
On the other hand the displayed trace can show the results of two consecutive
sweeps.
Device-specific, command or query
SYST:DATA:SIZE ALL
Select data transfer with full buffer size in order to view the trace data from a single
sweep..
SYSTem:DATE <Year>, <Month>, <Day>
Queries or defines the instrument's current date setting. The setting command requires administrator
rights.
Four-digit number
Two-digit number, 01 (for January) to 12 (for December)
Two-digit number, 01 to the number of days in the month
Confirmed, command or query
SYSTem:DATE?
Response: 2012,05,01 – it is the 1st of May, 2012.