R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 1060
Number of connected extension units R&S ZVAX-TRM.
– (*RST has no effect on external devices)
Device-specific, query only
SYST:COMM:RDEV:TEUN:COUN?
If a single R&S ZVAX-TRM is connected, the query returns 1.
SYSTem:COMMunicate:RDEVice:TEUNit:IDN?
Queries the device ID of a connected extension unit R&S ZVAX-TRM.
Manufacturer, type of the extension unit, 10-digit order number, serial number, see
example below. If no extension unit is connected, an execution error is generated.
– (*RST has no effect on external devices)
Device-specific, query only
SYST:COMM:RDEV:TEUN:IDN?
Query the device ID. A possible device response is Rohde&Schwarz,ED
INTERFACE ZVAX-TRM24,13226500K50100001, where 13221.6500.K24 is the
order number of the extension unit R&S ZVAX-TRM24, 100001 is the serial number
of the connected unit.
SYST:COMM:RDEV:TEUN:OPT?
Query the installed hardware options of the extension unit. A possible device
response is ZVAX-TRM24B112,ZVAX-TRM24B134,ZVAX-TRM24B213,ZVAX-
TRM24B224,ZVAX-TRM24B712,ZVAX24B73.
SYSTem:COMMunicate:RDEVice:TEUNit<zvax_trm_no>:OPT?
Queries the hardware options of a detected extension unit R&S ZVAX-TRM.
Comma-separated list of options, see example.
– (*RST has no effect on external devices)
Device-specific, query only
See SYSTem:COMMunicate:RDEVice:TEUNit:IDN?
SYSTem:CORRection:FMPort[:STATe] <Boolean>
Enables or disables Fast Multiport Correction. In manual control, the setting appears in the System
Configuration – General tab.
ON | OFF – Enable or disable fast multiport correction.