Operating Manual 1145.1084.12 – 30 1059
SYST:COMM:RDEV:PMET:DEL
Clear the power meter configuration table.
SYST:COMM:RDEV:PMET:DEF?
Query the power meter configuration. The analyzer returns an error message
because the power meter no. 1 is no longer configured.
SYSTem:COMMunicate:RDEVice:PMETer<pmeter_no>:DELete
Clears the configuration table for external power meters.
Number of the configured power meter. This number is ignored, the command
clears all entries in the configuration table.
Device-specific, no query
See SYSTem:COMMunicate:RDEVice:PMETer<pmeter_no>:DEFine
SYSTem:COMMunicate:RDEVice:RECeiver:DEFine '<IP_address>'
Configures a network analyzer as an external receiver for the mixer delay measurement and adds it to the
list of available receivers.
IP address (string parameter)
*RST has no effect on the configured receivers.
Device-specific, command or query
See [SENSe<Ch>:]FREQuency:MDELay:RECeiver
SYSTem:COMMunicate:RDEVice:RECeiver:DELete
Clears the configuration table for external receivers for the mixer delay measurement.
Device-specific, no query
See [SENSe<Ch>:]FREQuency:MDELay:RECeiver
SYSTem:COMMunicate:RDEVice:TEUNit:COUNt?
Queries the number of connected extension units R&S ZVAX-TRM.
Because the firmware currently only supports a single extension unit (R&S ZVAX-TRM or R&S ZVAX24),
the query returns
1, if the first detected extension unit is a R&S ZVAX-TRM and
0 otherwise