R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 912
ON | OFF – Enable / disable division
Device-specific, command or query.
[SENSe<Ch>:]FREQuency:MDELay:RECeiver INTernal | EXTernal
Selects the measurement setup/receiver configuration for the mixer delay measurement.
INTernal – internal receiver (one network analyzer only)
EXTernal – external receiver (second analyzer)
Device-specific, command or query.
[SENSe<Ch>:]FREQuency:MDELay:RECeiver:USE LAN1 | LAN2
Selects the LAN connection for the mixer delay measurement with external receiver. The command has
no effect if a single analyzer (internal receiver) is used.
LAN1 | LAN – LAN connector on the rear panel of the analyzer
Device-specific, command or query.
See [SENSe<Ch>:]FREQuency:MDELay:RECeiver
[SENSe<Ch>:]FREQuency:MDELay:UTONe PORT | GENerator, <source_no>
Selects the source for the upper tone signal that is used for the mixer delay measurement.
Upper tone source:
NONE – no source selected
PORT – internal source at port <source_no>
GENerator – external generator no. <source_no>, as defined in the System
Configuration – External Generators dialog.
Number of the port for the internal source or of the generator (to be omitted for
source: NONE)
1 to port number of the analyzer/number of external generators [–]