ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 911
option
- for network analyzers R&S ZVT20 that are equipped with option R&S ZVT20-B11
The internal combiner requires the following port configuration: Lower Tone: Port 1, Upper Tone: Port 3.
The two-tone signal is available at port 1.
ON | OFF – enables or disables the measurement using the internal
combiner
Device-specific, command or query.
See [SENSe<Ch>:]FREQuency:MDELay:APERture
[SENSe<Ch>:]FREQuency:MDELay:CONVersion OFF
Disables the mixer delay measurement and switches back to normal (non frequency-converting) mode.
Disable mixer delay measurement
Device-specific, no query.
See [SENSe<Ch>:]FREQuency:MDELay:APERture
[SENSe<Ch>:]FREQuency:MDELay:CORRection[:STATe] <Boolean>
Qualifies whether the analyzer applies the current mixer delay meas. calibration data.
The analyzer automatically applies the calibration data when a calibration file is loaded
(MMEMory:LOAD:MDCData) or after a calibration sweep has been completed
([SENSe<Ch>:]FREQuency:MDELay:ACQuire).
ON | OFF – Apply / ignore calibration data
Device-specific, command or query.
[SENSe<Ch>:]FREQuency:MDELay:DIVide <Boolean>
Enables or disables the division of loaded mixer delay calibration data by two.