ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 1062
SYSTem:DISPlay:COLor DBACkground | LBACkground | BWLStyles | BWSolid
Selects the color scheme for all diagram areas in the active setup.
Black and white line styles
–
*RST does not affect the color settings; see also description of the Preset
command.
Device-specific, command or query
SYST:DISP:COL LBAC
Select a light background, e.g. to generate color hardcopies.
SYSTem:DISPlay:UPDate <Boolean> | ONCE | FREeze
Switches the display on or off while the analyzer is in the remote state. The command has no effect while
the analyzer is in the Local operating state.
Switching off the display speeds up the measurement. This command may have an impact on the
update of trace and channel settings; see SYSTem:SETTings:UPDate.
ON | OFF – Switch the display on or off. If the display is switched on, the analyzer
shows the diagrams and traces like in manual control.
Switch the display on and show the current trace.
In continuous sweep mode the measurement is continued, however, the results
are not updated unless the setup has changed. In single sweep mode
SYSTem:DISPlay:UPDate ONCE must be called again in order to reflect any
changes in the trace setup.
This parameter can be used for occasional checks of the measurement results or
settings. Compared to the ON setting, ONCE does not slow down the measurement
speed.
Switch the display on, show the current trace and freeze the display, i.e. don't
update the display even if the setup changes.
*RST has no effect on the display update setting.
Device-specific, command or query
INITiate:CONTinuous ON
SYST:DISP:UPD ON
Turn on contiuous mode and switch the display on to view the traces and diagrams
DISPlay:UPDate ONCE