R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 1069
See SYSTem:PRESet:USER[:STATe]
SYSTem:PRIority NORMal | ANORmal | HIGH
Selects the priority of the running NWA application.
Normal priority, priority above normal, high priority
NORMal (*RST does not affect the priority settings)
Device-specific, command or query
SYST:PRI ANOR
Enhance the priority of the NWA application.
SYSTem:SETTings:UPDate ONCE
Initiates an immediate update of the channel or trace settings.
The command has an effect if the analyzer operates in single sweep mode
(INITiate<Ch>:CONTinuous OFF) and if the display update is switched off
(SYSTem:DISPlay:UPDate OFF). In this scenario, a change of the channel or trace settings is usually
not taken into account immediately. The analyzer aborts the current sweep sequence but postpones the
preparation of the next sweep sequence (including a re-configuration of all settings made during the last
sweep period) until the next single sweep sequence is initiated. Several settings can be made en bloc,
which generally saves time.
See also programming example in section Initiating Measurements, Speed Considerations.
SYSTem:SETTings:UPDate ONCE causes the analyzer to apply the settings immediately without waiting
for the end of the current single sweep sequence. The command has no effect in continuous sweep mode
or if the display update is switched on.
The settings are also updated when the continuous sweep mode is activated deliberately
(INITiate<Ch>:CONTinuous ON).
Causes an immediate update of the settings.
INIT:CONT OFF
Activate single sweep mode.
SYST:SETT:UPD ONCE
Update the settings made during the current single sweep period.