R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 623
Calculate the magnitude of z, to be displayed in a Cartesian diagram with a linear scale
Calculate the magnitude of z, displayed in a Cartesian diagram with a logarithmic scale
Phase of z, displayed in a Cartesian diagram with a linear vertical axis
Unwrapped phase of z, displayed in a Cartesian diagram with a linear vertical axis
Magnitude and phase, displayed in a polar diagram
Magnitude and phase, displayed in a Smith chart
Magnitude and phase, displayed in an inverted Smith chart
Group delay, displayed in a Cartesian diagram
Real part (x), displayed in a Cartesian diagram
Imaginary part (y), displayed in a Cartesian diagram
Standing Wave Ratio (SWR), displayed in a Cartesian diagram
x, y, displayed in a polar diagram
Magnitude (sqrt(x
2
+ y
2
)), displayed in a Cartesian diagram with a logarithmic scale
CALCulate<Chn>:FORMat:WQUType POWer | VOLTage
Selects the physical unit of the displayed trace.
Channel number used to identify the active trace
Confirmed, command or query
CALC4:PAR:SDEF 'Ch4Tr1', 'b1'
Create channel 4 and a trace named Ch4Tr1 to measure the wave quantity b
1
.
The trace becomes the active trace in channel 4.
CALC4:FORM:WQUT VOLT
Select voltage units for the created trace (identified by the suffix 4).
CALCulate<Chn>:FSIMulator...
This subsystem enables the virtual SAW Matching Network and defines its parameters. When the virtual
network is enabled, port 1 serves as an unbalanced port, the physical ports no. 2 and 4 of the analyzer are
combined to a logical (balanced) port no. 2.
The commands in this subsystem are supported for the sake of compatibility with older firmware
versions. Starting with firmware V1.70 the commands have been superseded by the
CALCulate:TRANsform:VNETwork... subsystem. Virtual network (de-)embedding is controlled by the
Virtual Transform menu.