R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 706
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding:TNDefinition
FIMPort | SL | SC
Selects the circuit model for ground loop deembedding.
Possible circuit models (character data); see circuit models.
Device-specific, command or query.
See CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding[:STATe]
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding:PARameters:C CSL |
LSC | CSC | SLC | SCL | SCC,<capacitance>
Specifies the capacitance values C1, C2 in the different circuit models for ground loop embedding.
SC – Shunt C circuit model (character data); see circuit models.
Capacitance C for ground loop embedding.
–1mF to 1 mF [F]. The increment is 1 fF (10
–15
F).
Device-specific, command or query. In the query form, only the first parameter
is required; the <capacitance> must be omitted.
*RST; :CALC:TRAN:VNET:GLO:EMB:PAR:C? SC
Query the default capacitance for ground loop embedding. The response is
1E-012 (1 pF).
CALC:TRAN:VNET:GLO:EMB:PAR:C SC, 2.2E-12
Increase the capacitance to 2.2 pF.
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding:PARameters:L SL,
<inductance>
Specifies the inductance value for ground loop embedding.
SL – Shunt L circuit model (character data); see circuit models.
Inductance L for ground loop embedding.
–1H to 1 H [H]. The increment is 1 pH (10
–12
H).
Device-specific, command or query. In the query form, only the first parameter
is required; the <inductance> must be omitted.