R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 716
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<Ph_pt>:PARame
ters:C<no> CSL | LSC | CSC | SLC | SCL | SCC,<numeric_value>
Specifies the capacitance values C1, C2 in the different circuit models for single ended port deembedding.
Number of capacitance in circuit model. If unspecified the numeric suffix is set
to 1.
1 to 2 [–] The number of capacitances depends on the selected circuit model.
Possible circuit models (character data); see circuit models.
– (see command type description below)
Capacitance C<no> for the specified circuit model.
–1mF to 1 mF [F]. The increment is 1 fF (10
–15
F).
Device-specific, command or query. In the query form, the <numeric_value>
parameter must be omitted. The command returns the capacitance value for
the specified circuit model.
*RST; :CALC:TRAN:VNET:SEND:DEEM:PAR:C2? CSC
Query the default capacitance C2 for the Serial C, shunt C circuit model. The
response is 1E-012 (1 pF).
CALC:TRAN:VNET:SEND:DEEM:PAR:C2 CSC, 2.2E-12
Increase the capacitance to 2.2 pF.
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<Ph_pt>:PARame
ters:L<no> CSL | LSC | LSL | SLC | SCL | SLL,<numeric_value>
Specifies the inductance values L1, L2 in the different circuit models for single ended port deembedding.
Number of inductance in circuit model. If unspecified the numeric suffix is set
to 1.
1 to 2 [–] The number of inductances depends on the selected circuit model.
Possible circuit models (character data); see circuit models.
– (see command type description below)
Inductance L<no> for the specified circuit model.
–1H to 1 H [H]. The increment is 1 pH (10
–12
H).
Device-specific, command or query. In the query form, the <numeric_value>
parameter must be omitted. The command returns the inductance value for the