R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 771
MMEM:LOAD:CKIT:SDAT 'N 50 Ohm','Default Kit',MOPEN,'Test
data','test.s1p',1
Load the file Test.s1p from the current directory in order
to define the properties of an Open (m) standard in the
cal kit named Default Kit for the N 50 Ω connector type.
Assign the label Test data and specify that the standard
data is valid for port no. 1 only.
Connector and cal kit naming conventions
Connector and calibration kit names must be entered as string parameters. The strings contain the
connector and cal kit names used in the Calibration Kits dialog; a Ω in the name must be replaced by
'Ohm', e.g.:
'NewKit1' denotes the user-defined calibration kit NewKit1.
'N 50 Ohm Ideal Kit' denotes the N 50 Ω Ideal Kit.
'ZV-Z21 typical' denotes the cal kit ZV-Z21 typical.
MMEMory:LOAD:CKIT:UDIRectory '<directory>'
Specifies the Directory for Additionally Available Cal Kits and Conn Types. All cal kit files in the specified
directory will be (re-)loaded automatically as predefined kits (i.e. read-only kits which cannot be modified)
every time the NWA application is started.
String parameter to specify the directory path. The directory can be created
separately (MMEMory:MDIRectory). An empty string means that no cal kit files
will be loaded.
Device-specific, no query.
MMEM:LOAD:CKIT:UDIR 'C:\Program Files\Rohde&Schwarz\Network
Analyzer\Calibration\Kits\Autoload'
Specifiy the directory for additionally available cal kits.
MMEM:MDIR 'C:\Program Files\Rohde&Schwarz\Network
Analyzer\Calibration\Kits\Autoload'
Create the specified 'Autoload' directory.
MMEM:STOR:CKIT 'New_kit', 'C:\Program
Files\Rohde&Schwarz\Network
Analyzer\Calibration\Kits\New_kit.calkit'
Store the data for the user-defined cal kit New_kit to the
'Autoload' directory.