[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe <Boolean>
Activates or deactivates the calibration mode where the raw measurement data of the standards is stored
after the calibration is completed. The setting is valid for the current calibration, where it overwrites the
global setting ([SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe:DEFault). A new
calibration deletes the calibration data acquired in previous calibrations.
ON | OFF - Keep measurement data on or off.
Device-specific, command or query
CORR:COLL:RSAV:DEF ON
Generally keep the raw measurement data of the standards after the
calibration is completed.
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard as calibration
type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the raw measurement
results of this standard.
CORR:COLL:RSAV OFF
To save disk space, delete the current raw calibration data after the calibration
is completed.
CORR:COLL:SAVE:SEL
Calculate the system error correction data and apply it to the active channel.
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe:DEFault <Boolean>
Activates or deactivates the calibration mode where the raw measurement data of the standards is stored
after the calibration is completed. The setting remains valid for all subsequent calibrations until it is
changed explicitly. A new calibration deletes the calibration data acquired in previous calibrations.
ON | OFF - Keep measurement data on or off.
Device-specific, command or query
See [SENSe<Ch>:]CORRection:COLLect[:AQUire]:RSAVe.