CORR:COLL:AUTO:ASSignment1:DEFine?
Query the first port assignment. The response is 1, 1, 2, 2.
--> Connect the calibration unit between port 1 and 2 of the analyzer.
CORR:COLL:AUTO:ASSignment1:ACQuire
Acquire the calibration data for the first port assignment.
--> Disconnect the analyzer port 1, connect port 3 instead.
CORR:COLL:AUTO:ASSignment2:ACQuire
Acquire the calibration data for the second port assignment.
--> Disconnect the analyzer port 2, connect port 4 instead.
CORR:COLL:AUTO:ASSignment3:ACQuire
Acquire the calibration data for the third port assignment.
CORR:COLL:AUTO:SAVE
Calculate the error terms from all three calibrations, store them, and apply
the calibration to the active channel.
CORR:COLL:AUTO:ASSignment:DELete:ALL
Delete all port assignments.
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DELete:ALL
Deletes all port assignments.
Current number of the port assignment. This suffix is ignored; the command deletes
all assignments.
Device-specific, no query
See [SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine
[SENSe<Ch>:]CORRection:COLLect:AUTO:CKIT
'<file_name>'[,<characterizationMethod>]
Generates a cal kit file with the specified name containing the cal kit data of the active calibration unit
(SYSTem:COMMunicate:RDEVice:AKAL:ADDRess). The cal kit file can be stored in the calibration unit