SCPI Command Reference
Operating Manual 1145.1084.12 – 30 837
or written to a directory on the analyzer.
The optional <characterizationMethod> defines the calibration types to be supported with this
characterization (implicitly: the standards to be included in the characterization).
Note that this command assumes the "canonical" assignment of cal unit ports and physical VNA ports:
VNA port 1 assigned to cal unit port 1, VNA port 2 assigned to cal unit port 2 etc. The command
[SENSe:]CORRection:COLLect:AUTO:CKIT:PORTs allows for flexible port assignments.
Name and (possibly) directory of the created characterization / cal kit
file (string parameter, extension *.calkit):
If no path and directory is specified, the file is stored in the
calibration unit (see also
[SENSe<Ch>:]CORRection:COLLect:AUTO). The factory
calibration data on the unit is not overwritten. If the file name is
prefixed by SD:, the file is saved to the SD card inserted at the
calibration unit (if available/accessible).
If a path and directory is specified, the file is stored in the
directory on the analyzer's hard disk.
The default directory (MMEMory:CDIRectory) is not used.
The calibration types that shall be supported with this characterization:
FOPort – Full One Port (one port standards only)
FNPort – Full N Port (one and two port standards)
DEFault [or omitted] – Default characterization method of the active cal
unit (property [CAL_PROCEDURE] > ExtCalProc set in the calu's
config.ini)
Device-specific, no query
CORR:COLL:AUTO:CKIT 'AutoCalChar.calkit'
Generate a cal kit file AutoCalChar.calkit for the active calibration unit
(supporting the CalUnit's default calibration procedure) and store it
internally.
[SENSe<Ch>:]CORRection:COLLect:AUTO:CKIT:PORTs
'<file_name>',<characterizationMethod>,
<VNA_port_1>,<CalU_port_1>,<VNA_port_2>,<CalU_port_2>...
Generates a cal kit file with the specified name containing the cal kit data of the active calibration unit
(SYSTem:COMMunicate:RDEVice:AKAL:ADDRess). The cal kit file can be stored in the calibration unit
or written to a directory on the analyzer.
The <characterizationMethod> defines the calibration types to be supported with this characterization
(implicitly: the standards to be included in the characterization). The port pairs
<VNA_port_i>,<CalU_port_i> specify the current port connections between VNA and CalU.
Similar logic as [SENSe:]CORRection:COLLect:AUTO:CKIT, but with flexible port assignment.