[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
Initiates an automatic calibration for a previously defined port assignment numbered <Asg>.
A complete, valid set of port assignments must be defined before you can initiate a calibration; see
program example.
Current number of the port assignment.
Device-specific, no query
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine
<analyzer_port_no>,<cal_unit_port_no>{,<analyzer_port_no>,<cal_unit_port_no>}
Defines a port assignment numbered <Asg>. This command is particularly relevant for multiple port
assignments.
Current number of the port assignment
Port numbers of the analyzer. For an n-port automatic calibration (n = 1 to
4), n arbitrary (not necessarily consecutive) port numbers must be specified.
Port numbers of the cal unit. The rules for valid port assignments depend on
the calibration type selected via
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure; see section
Using Multiple Port Assignments. Only port combinations with assigned cal
unit ports must be specified; see example.
Device-specific, command or query
Suppose that a 2-port calibration unit with a factory characterization is
connected to a four-port analyzer.
CORR:COLL:AUTO:CONFigure FNPort, ' '
Select the factory cal kit file (empty string) and a full n-port (i.e. 4-port)
calibration.
CORR:COLL:AUTO:ASSignment1:DEFine 1, 1, 2, 2
CORR:COLL:AUTO:ASSignment2:DEFine 2, 2, 3, 1
CORR:COLL:AUTO:ASSignment3:DEFine 3, 1, 4, 2
Define three port assignments, corresponding to the following table in the
Calibration Settings dialog.