R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 869
Device-specific, command or query
See Noise Figure Measurement (condensed programming examples)
[SENSe<Chn>:]CORRection:NSTate?
Returns the noise figure calibration state label of active trace in channel <Chn>. The active trace must be
a noise figure trace, otherwise an empty string is returned.
Channel number used to identify the active trace.
String variable containing the calibration state label in the trace list ('NCal', 'NCai',
'NCax', 'NCa?', 'NCal Inv', 'NCal Off'; see Noise Figure Calibration State Labels).
'' (empty string, is returned e.g. if no noise figure calibration is available for trace
<Chn>)
Device-specific, query only
See Noise Figure Measurement (condensed programming examples)
[SENSe<Ch>:]CORRection:OFFSet<port_no>:DFComp[:STATe]? <Boolean>
Returns whether a direct fixture compensation has been carried out at port no. <port_no>. A direct fixture
compensation resets the offset parameters to zero, the analyzer uses calculated transmission factors
instead.
Channel number of the offset-corrected channel
Port number of the analyzer.
1 – direct fixture compensation data used
0 – no direct fixture compensation data used
Device-specific, query only
*RST; :CORR:OFFS:DFC?
Reset the instrument and query whether the analyzer uses direct fixture
compensation data at port 1. The response is 0.
[SENSe<Ch>:]CORRection:OFFSet<port_no>:MAGNitude <DC_loss>
[Deprecated command]
Defines the frequency-independent part (DC value) of the offset loss.