Use the [SENSe<Ch>:]CORRection:LOSS<port_no>... commands to define the complete set of
loss offset parameters. [SENSe<Ch>:]CORRection:OFFSet<port_no>:MAGNitude is equivalent to
[SENSe<Ch>:]CORRection:LOSS<port_no>:OFFSet
Channel number of the offset-corrected channel
Port number of the analyzer
Frequency-independent part of the offset loss
-200 dB to +200 dB [dB]. The increment (UP/DOWN) is 0.001 dB.
Confirmed, command or query
[SENSe<Ch>:]CORRection:OFFSet<port_no>[:STATe] <Boolean>
Resets the offset parameters for all test ports to zero or queries whether any of the offset parameters are
different from zero.
Channel number of the offset-corrected channel
Port number of the analyzer. This numeric suffix is ignored; the command
affects the parameters of all ports.
The parameter function depends on whether the command is used as a setting
command or as a query:
For setting command:
ON – no effect
OFF – resets all length offsets to zero
For query:
1 – at least one length offset is different from zero
0 – all length offsets are zero
Device-specific, command or query
*RST; :CORR:OFFS?
Reset the instrument and query whether the length offset parameters have
been reset as well. The response is 0.
SENSe<Ch>:CORRection:POWer
[SENSe<Ch>:]CORRection:POWer<port_no>...
This subsystem controls the receiver power calibration.
[SENSe<Ch>:]CORRection:POWer<port_no>:ACQuire AWAVe | BWAVe | B1 | B2 |
B3 | B4[,<cal_port>, PORT | GENerator | CONVerter, <source_no>, AWAVe |