[SENSe<Ch>:]FREQuency<Pt>:OFFSet:PWAVes <offset>
Defines the constant receiver frequency offset for primed wave quantities relative to the common receiver
frequency.
Test port number of the analyzer.
Depending on the instrument model [Hz]. The increment (parameters UP or
DOWN) is 1 kHz.
Device-specific, command or query.
FREQ:STAR 2GHz; STOP 3 GHz
Define a sweep range between 2 GHz and 3 GHz.
FREQ:OFFS:PWAV 10kHz; WAV 20kHz
Modify the receiver frequency settings.
[SENSe<Ch>:]FREQuency<Pt>:OFFSet:WAVes <offset>
Defines the constant receiver frequency offset for unprimed wave quantities relative to the common
receiver frequency.
The offset for unprimed waves is not displayed in the Port Configuration dialog.
Test port number of the analyzer.
Depending on the instrument model [Hz]. The increment (parameters UP or
DOWN) is 1 kHz.
Device-specific, command or query.
See [SENSe<Ch>:]FREQuency<Pt>:OFFSet:PWAVes
[SENSe<Ch>:]FREQuency:SBANd POSitive | NEGative | AUTO
Defines whether the analyzer measures with a local oscillator frequency LO below or above the RF input
frequency.
In a segmented frequency sweep, it is possible to set the sideband (SBANd) parameter individually for
each segment; see [SENSe<Ch>:]SEGMent<Seg>:DEFine. The [SENSe<Ch>:]FREQuency:SBANd
setting is global and not valid for segmented sweeps. The two sideband settings do not overwrite each
other.