R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 917
Device-specific, command or query
See Noise Figure Measurement (condensed programming examples)
[SENSe<Ch>:]NFIGure:RFICorr <Boolean>
Enables or disables the RF image correction for mixer noise figure measurements.
ON | OFF – Enable or disable RF image correction.
Device-specific, command or query
See Noise Figure Measurement (condensed programming examples)
[SENSe<Ch>:]NFIGure:SEQuential <Boolean>
Selects sequential or simultaneous measurement mode.
Noise figure measurements for mixers with embedded LO
([SENSe<Ch>:]FREQuency:CONVersion:MIXer:LOPort<Stg> EMBedded) cannot be performed in
simultaneous mode.
ON | OFF – Sequential or simultaneous measurement
Device-specific, command or query
See Noise Figure Measurement (condensed programming examples)
[SENSe<Chn>:]PAE...
This subsystem configures the measurement of the Power Added Efficiency (PAE).
[SENSe<Chn>:]PAE:C <numeric_value>
Defines the constant c for the PAE measurement.
Channel number used to identify the active trace.
Value of the constant c in the PAE test model
([SENSe<Chn>:]PAE:EXPRession). The value is used if one of the test
models C10 | C1 | CK11 is selected.