R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 956
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl:CONTrol <Boolean>
Selects common or independent Meas. Delay settings for the sweep segments.
Sweep segment number. This suffix is ignored; the setting controls the whole
segmented sweep.
ON: Use independent delay settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl.
OFF: Reset the delay in all sweep segments to the delay for unsegmented sweeps,
defined via [SENSe<Ch>:]SWEep:DWELl. ON will not restore the previous values.
OFF. The parameter is automatically switched to ON when a delay is entered using
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl.
Device-specific, command or query.
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings and thus
0 s meas. delay.
SEGM:SWE:DWELl 0.1
Increase the meas. delay to 0.1 s.
SEGM:SWE:DWELl:CONT OFF
Couple the meas. delay in all segments and reset the delay in segment no. 1 to the
initial value of 0 s.
[SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts <Points>
Defines the total number of measurement Points in sweep segment no. <Seg>.
Number of points in the segment.
1 to 60001. [1] 1 is allowed if start and stop frequencies are equal.
Device-specific, command or query
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default
settings.
SEGM:SWE:POIN 401
Increase the number of points to 401.