EasyManua.ls Logo

Rohde & Schwarz FSVA - Page 947

Rohde & Schwarz FSVA
1106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSVA/FSV
947Operating Manual 1176.7510.02 ─ 09
MMEMory:SELect[:ITEM]:HWSettings <State>
This command includes the hardware settings in the settings file to be stored/loaded.
The hardware settings include:
current configuration of general device parameters (general setup)
current setting of the measurement hardware including markers
activated limit lines:
A settings file may include 8 limit lines at maximum. This number includes the activated
limit lines and, if available, the de-activated limit lines last used.
Therefore the combination of the non-activated restored limit lines depends on the
sequence of use with the MMEMory:LOAD:STATe on page 942 command.
user-defined color setting
configuration for the hardcopy output
Parameters:
<State> ON | OFF
*RST: ON
Example:
MMEM:SEL:HWS ON
Manual operation: See "Select Items" on page 198
MMEMory:SELect[:ITEM]:LINes:ALL <State>
This command includes all limit lines (activated and de-activated) in the settings file to
be stored/loaded.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
MMEM:SEL:LIN:ALL ON
Manual operation: See "Select Items" on page 198
MMEMory:SELect[:ITEM]:NONE
This command excludes all items from the settings file to be stored/loaded.
Example:
MMEM:SEL:NONE
Usage: Event
Manual operation: See "Disable all Items" on page 199
MMEMory:SELect[:ITEM]:SGRam <boolean>
This command adds the spectrogram data to the list of data subsets of a save/recall
device setting.
Remote Control – Commands

Table of Contents