Remote control commands
R&S
®
SMB100B
401User Manual 1178.3711.02 ─ 08
Setting parameters:
<IP> string
*RST: 0
Example: See Example"Detecting and assigning a power sensor"
on page 398.
Usage: Setting only
Manual operation: See "Add LAN Sensor settings" on page 171
:SLISt:SCAN:USENsor
<DeviceID>, <Serial>
Scans for R&S NRP power sensors connected over a USB interface.
Parameters:
<Serial> integer
Range: 0 to 999999
Setting parameters:
<DeviceID> String or Integer
Range: 0 to 999999
*RST: 0
Example: See Example"Detecting and assigning a power sensor"
on page 398.
Usage: Setting only
Manual operation: See "Add USB Sensor settings" on page 171
:SLISt:CLEar:LAN
Removes all R&S NRP power sensors connected in the LAN from the list.
Example:
See Example"Detecting and assigning a power sensor"
on page 398.
Usage: Event
Manual operation: See "Clear" on page 171
:SLISt:CLEar:USB
Removes all R&S NRP power sensors connected over USB from the list.
Example:
See Example"Detecting and assigning a power sensor"
on page 398.
Usage: Event
Manual operation: See "Clear" on page 171
SENSe, READ, INITiate and SLISt subsystems