Remote control commands
R&S
®
SMB100B
431User Manual 1178.3711.02 ─ 08
[:SOURce<hw>]:PULM:DOUBle:STATe <State>
Provided for backward compatibility with former Rohde & Schwarz signal generators.
Works like the command [:SOURce<hw>]:PULM:MODEDOUBle.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
[:SOURce<hw>]:PULM:STATe
<State>
Activates pulse modulation.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example: See Example"Perform pulse modulation" on page 427.
Manual operation: See "State" on page 72
[:SOURce<hw>]:PULM:SOURce <Source>
Selects between the internal (pulse generator) or an external pulse signal for the mod-
ulation.
Parameters:
<Source> INTernal | EXTernal
*RST: INTernal
Example: See Example"Perform pulse modulation" on page 427.
Manual operation: See "Source" on page 72
[:SOURce<hw>]:PULM:TTYPe <Source>
Sets the transition mode for the pulse signal.
Parameters:
<Source> SMOothed | FAST
SMOothed
flattens the slew rate, resulting in longer rise/fall times.
FAST
enables fast transitions with shortest rise and fall times.
*RST: FAST
Example: See Example"Perform pulse modulation" on page 427.
Manual operation: See "Transition Type" on page 72
SOURce subsystem