Remote control commands
R&S
®
SMB100B
428User Manual 1178.3711.02 ─ 08
// Alternatively configure the pulse modulation settings for
// external modulation source
// Select the source, set the polarity of the external signal,
// select the impedance for the external pulse modulation input/
// for the external pulse modulation trigger input
SOURce:PULM:SOURce EXT
SOURce:PULM:POLarity NORMal
SOURce:PULM:IMPedance G1K
// Configure the pulse generator settings
// Set pulse period, width, and delay
SOURce:PULM:PERiod 10 us
SOURce:PULM:WIDth 8 us
SOURce:PULM:DOUBle:WIDTh 0.0000012
SOURce:PULM:DOUBle:DELay 0.0000045
// Route the signal to the signal valid output connector
// to use the signal for synchronizing the pulse modulation.
:SOURCe:PULM:OUTPut:SYNC ON
// Activate the signal output
SOURce:PGENerator:OUTPut:STATe 1
SOURce:PULM:STATe 1
OUTPut1:STATe 1
The following commands are available:
[:SOURce<hw>]:PULM:MODE........................................................................................ 428
[:SOURce<hw>]:PULM:TRIGger:MODE...........................................................................429
[:SOURce]:PULM[:INTernal][:TRAin]:TRIGger:IMMediate...................................................429
[:SOURce<hw>]:PULM:PERiod....................................................................................... 429
[:SOURce<hw>]:PULM:WIDTh........................................................................................430
[:SOURce<hw>]:PULM:DELay........................................................................................ 430
[:SOURce<hw>]:PULM:DOUBle:DELay............................................................................430
[:SOURce<hw>]:PULM:DOUBle:WIDTh........................................................................... 430
[:SOURce<hw>]:PULM:DOUBle:STATe............................................................................431
[:SOURce<hw>]:PULM:STATe.........................................................................................431
[:SOURce<hw>]:PULM:SOURce..................................................................................... 431
[:SOURce<hw>]:PULM:TTYPe........................................................................................431
[:SOURce<hw>]:PULM:OUTPut:SYNC[:STATe].................................................................432
[:SOURce<hw>]:PULM:POLarity..................................................................................... 432
[:SOURce<hw>]:PULM:IMPedance..................................................................................432
[:SOURce<hw>]:PULM:MODE <Mode>
Selects the mode for the pulse modulation.
Parameters:
<Mode> SINGle | DOUBle | PTRain
SINGle
Generates a single pulse.