Remote control commands
R&S
®
SMB100B
429User Manual 1178.3711.02 ─ 08
DOUBle
Generates two pulses within one pulse period.
PTRain
Generates a user-defined pulse train.
Specify the pulse sequence with the commands:
[:SOURce<hw>]:PULM:TRAin:ONTime
[:SOURce<hw>]:PULM:TRAin:OFFTime
[:SOURce<hw>]:PULM:TRAin:REPetition
*RST: SINGle
Example: See Example"Perform pulse modulation" on page 427.
Options: PTRain requires R&S SMBB-K27
Manual operation: See "Pulse Mode" on page 88
[:SOURce<hw>]:PULM:TRIGger:MODE <Mode>
Selects a trigger mode - auto, single, external, external single or external gated - for
generating the modulation signal.
Parameters:
<Mode> AUTO | EXTernal | EGATe | SINGle | ESINgle
*RST: AUTO
Example: See Example"Perform pulse modulation" on page 427.
Manual operation: See "Trigger Mode" on page 89
[:SOURce]:PULM[:INTernal][:TRAin]:TRIGger:IMMediate
If [:SOURce<hw>]:PULM:TRIGger:MODESINGle, triggers the pulse generator.
Example:
See Example"Perform pulse modulation" on page 427.
Manual operation: See "Execute Single Trigger" on page 90
[:SOURce<hw>]:PULM:PERiod <Period>
Sets the period of the generated pulse, that means the repetition frequency of the inter-
nally generated modulation signal.
Parameters:
<Period> float
The minimum value depends on the installed options
R&S SMBB-K22 or R&S SMBB-K23
Range: 20E-9 to 100
Increment: 5E-9
*RST: 10E-6
Example: See Example"Perform pulse modulation" on page 427.