Analog modulations
R&S
®
SMB100B
89User Manual 1178.3711.02 ─ 08
"Train"
Option: R&S SMBB-K27
Generates a user-defined pulse train.
1 = 1st pulse: On-time
2 = 1st pulse: Off-time
3 = 1st pulse: pulse period
4 = 2nd pulse: repeated 3 times
5 = 3rd pulse
6 = 4th pulse
See Chapter 5.4.4.2, "Pulse generator > pulse train settings",
on page 91.
Remote command:
[:SOURce<hw>]:PULM:MODE on page 428
Trigger Mode
Selects between continuous triggering or triggering initiated by a trigger event from an
external signal.
See also "Input and output connectors" on page 69.
"Auto"
Generates the internal modulation signal continuously.
"Single"
The pulse generator is triggered by an internal trigger event, initiated
with Execute Single Trigger.
The pulse signal is generated once.
"Ext Single"
Generates a single pulse signal, triggered by an external signal.
"Ext Triggered"
Generates the pulse signal each time an external trigger event
occurs.
"Ext Gated"
Generates the signal triggered by an external gate signal.
Remote command:
[:SOURce<hw>]:PULM:TRIGger:MODE on page 429
Pulse Period
Sets the repetition rate of the generated pulse signal.
Remote command:
[:SOURce<hw>]:PULM:PERiod on page 429