Remote control commands
R&S
®
SMB100B
435User Manual 1178.3711.02 ─ 08
[:SOURce<hw>]:PULM:TRAin:REPetition:POINts?
[:SOURce<hw>]:PULM:TRAin:ONTime:POINts?
[:SOURce<hw>]:PULM:TRAin:OFFTime:POINts?
Queries the number of on and off time entries and repetitions in the selected list.
Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example: See Example"Generating a pulse train signal" on page 432.
Usage: Query only
[:SOURce<hw>]:PULM:TRAin:REPetition <Repetition>
Sets the number of repetitions for each pulse on/off time value pair.
Parameters:
<Repetition> Repetition#1{, Repetition#2, ...}
0 = ignore value pair
Set "Repetition = 0" to skip a particular pulse without deleting
the pulse on/off time value pair
Range: 0 to 65535
Example: See Example"Generating a pulse train signal" on page 432.
Manual operation: See "Edit Pulse Train Data" on page 93
[:SOURce<hw>]:PULM:TRAin:SELect <Filename>
Selects or creates a data list in pulse train mode.
If the list with the selected name does not exist, a new list is created.
Parameters:
<Filename> string
Filename or complete file path; file extension can be omitted.
Example: See Example"Generating a pulse train signal" on page 432.
Manual operation: See "Pulse Train Data" on page 92
See "File Name" on page 97
See "Functions for handling of data lists" on page 208
Pulse train data exchange
Option: R&S SMBB-K27
With the following commands, you can configure pulse trains in ASCII format and
export or import them accordingly.