EasyManua.ls Logo

Rohde & Schwarz R&S SMB100B - Page 432

Rohde & Schwarz R&S SMB100B
641 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMB100B
432User Manual 1178.3711.02 ─ 08
[:SOURce<hw>]:PULM:OUTPut:SYNC[:STATe] <Sync>
Routes the pulse modulation signal to the [Signal Valid] connector.
Parameters:
<Sync> 1 | ON | 0 | OFF
*RST: OFF
Options: R&S SMBB-K23 (Pulse Generator)
Manual operation: See "Pulse Sync Out" on page 90
[:SOURce<hw>]:PULM:POLarity
<Polarity>
Sets the polarity of the externally applied modulation signal.
Parameters:
<Polarity> NORMal | INVerted
NORMal
Suppresses the RF signal during the pulse pause.
INVerted
Suppresses the RF signal during the pulse.
*RST: NORMal
Example: See Example"Perform pulse modulation" on page 427.
Manual operation: See "Polarity" on page 98
[:SOURce<hw>]:PULM:IMPedance <Impedance>
Sets the impedance for the external pulse trigger and pulse modulation input.
Parameters:
<Impedance> G50 | G10K
*RST: G10K
Example: See Example"Perform pulse modulation" on page 427.
Manual operation: See "Impedance" on page 98
Pulse train settings
Option: R&S SMBB-K27
This section describes the commands for the pulse train modulation and the associ-
ated file handling.
See Chapter 5.4.4.2, "Pulse generator > pulse train settings", on page 91.
Example: Generating a pulse train signal
The following settings are required to perform pulse train modulation.
SOURce subsystem