Remote control commands
R&S
®
SMB100B
499User Manual 1178.3711.02 ─ 08
●
OFFSet is set with [:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
Parameters:
<Amplitude> float
The following settings influence the value range:
OFFSet set with the command [:SOURce<hw>]:POWer[:
LEVel][:IMMediate]:OFFSet
Numerical value
Sets the level
UP|DOWN
Varies the level step by step.
The level is increased or decreased by the value set with the
command [:SOURce<hw>]:POWer:STEP[:INCRement].
Range: (Level_min + OFFSet) to (Level_max + OFFStet)
*RST: -30
Default unit: dBm
Example:
POWer -30
Sets the RF level
Example: See also [:SOURce<hw>]:POWer:STEP:MODE on page 497.
Manual operation: See "Amplitude" on page 64
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:REFLevel <ReferenceLevel>
Queries the reference level of the user correction.
The reference level is the sum of the amplitude and the level offset, set with the com-
mands [:SOURce<hw>]:POWer:POWer [:SOURce<hw>]:POWer[:LEVel][:
IMMediate]:OFFSet.
Parameters:
<ReferenceLevel> float
Range: -245 to 120
Increment: 0.01
*RST: -30
Manual operation: See "Used SMBB Settings For Measurement" on page 165
[:SOURce<hw>]:POWer:RANGe:LOWer?
[:SOURce<hw>]:POWer:RANGe:UPPer?
Queries the current interruption-free range of the level.
Return values:
<Upper> float
Increment: 0.01
Default unit: dBm
SOURce subsystem