Remote control commands
R&S
®
SMB100B
515User Manual 1178.3711.02 ─ 08
Parameters:
<CombPowStop> float
Range: -245 to 120
Increment: 0.01
*RST: -10
Example: See Example"Setup an RF frequency or power sweep"
on page 506.
Manual operation: See "Start Level / Stop Level" on page 137
[:SOURce<hw>]:SWEep:COMBined:COUNt
<StepCount>
Defines the number of sweeps you want to execute.
This parameter applies to [:SOURce<hw>]:SWEep:COMBined:MODE > SINGle. To
start the sweep signal generation, use the command [:SOURce<hw>]:SWEep:
COMBined:EXECute on page 515.
Parameters:
<StepCount> integer
Range: 1 to SeMAX_INT_STEP-1
*RST: 501
Example: See Example"Setup an RF frequency or power sweep"
on page 506.
Manual operation: See "Step Count" on page 137
[:SOURce<hw>]:SWEep:COMBined:DWELl <Dwell>
Sets the dwell time for the combined frequency / level sweep.
Parameters:
<Dwell> float
Range: 0.01 to 100
Increment: 100E-6
*RST: 0.01
Example: See Example"Setup an RF frequency or power sweep"
on page 506.
Manual operation: See "Dwell Time" on page 133
[:SOURce<hw>]:SWEep:COMBined:EXECute
Executes an RF frequency / level sweep cycle.
The command triggers one single sweep manually. Therefore, you can use it in manual
sweep mode, selected with the command [:SOURce<hw>]:SWEep:COMBined:MODE
> MANual.
SOURce subsystem