Remote control commands
R&S
®
SMB100B
528User Manual 1178.3711.02 ─ 08
Setting parameters:
<SecPassWord> string
Current security password
The default password is 123456.
Example:
SYSTem:SECurity:VOLMode:STATe "123456", 1
SYSTem:REBoot
Manual operation: See "Volatile Mode" on page 252
:SYSTem:COMMunicate:GPIB:LTERminator
<LTerminator>
Sets the terminator recognition for remote control via GPIB interface.
Parameters:
<LTerminator> STANdard | EOI
EOI
Recognizes an LF (Line Feed) as the terminator only when it is
sent with the line message EOI (End of Line). This setting is rec-
ommended particularly for binary block transmissions, as binary
blocks may coincidentally contain a characater with value LF
(Line Feed), although it is not determined as a terminator.
STANdard
Recognizes an LF (Line Feed) as the terminator regardless of
whether it is sent with or without EOI.
*RST: n.a. (factory preset: STANdard)
Example: See Example"Finding out VISA resource strings" on page 518.
:SYSTem:COMMunicate:GPIB:RESource?
Queries the visa resource string for remote control via the GPIB interface.
To change the GPIB address, use the command :SYSTem:COMMunicate:GPIB[:
SELF]:ADDRess.
Return values:
<Resource> string
Example: See Example"Finding out VISA resource strings" on page 518.
Usage: Query only
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <Address>
Sets the GPIB address.
Parameters:
<Address> integer
Range: 0 to 30
*RST: n.a. (factory preset: 28)
SYSTem subsystem