Remote control commands
R&S
®
SMB100B
538User Manual 1178.3711.02 ─ 08
:SYSTem:SRData?
Queris the SCPI recording data from the internal file.
This feature enables you to transfer an instrument configuration to other test environ-
ments, as e.g. laboratory virtual instruments.
Return values:
<FileData> block data
Example:
SYSTem:SRData?
// #3118:SOURce1:ROSCillator:SOURce EXT
// :SOURce1:FREQuency:CW 4000000000
// :SOURce1:FREQuency:OFFSet 1000000
// :SOURce1:AM1:STATe 1
// :OUTPut1:STATe 1
Usage: Query only
:SYSTem:STARtup:COMPlete?
Queries if the startup of the instrument is completed.
Return values:
<Complete> 1 | ON | 0 | OFF
*RST: 0
Example:
SYST:STAR:COMP?
Response: 1
the startup of the instrument is completed.
Usage: Query only
:SYSTem:DATE <Year>, <Month>, <Day>
Queries or sets the date for the instrument-internal calendar.
This is a password-protected function. Unlock the protection level 1 to access it.
Parameters:
<Year> integer
<Month> integer
Range: 1 to 12
<Day> integer
Range: 1 to 31
Example: See Example"Configuring date and time" on page 519.
Manual operation: See "Date" on page 593
SYSTem subsystem