Remote control commands
R&S
®
SMB100B
542User Manual 1178.3711.02 ─ 08
Return values:
<DeviceFootprint> string
Information on the instrument type, device identification and
details on the installed FW version, hardware and software
options.
Example:
:SYSTem:DFPR?
Usage: Query only
:SYSTem:REBoot
Reboots the instrument including the operating system.
Usage:
Event
:SYSTem:RESTart
Restarts the instrument without restarting the operating system.
Usage:
Event
:SYSTem:SHUTdown
Shuts down the instrument.
Usage:
Event
Manual operation: See "Shut down" on page 599
:SYSTem:WAIT <TimeMs>
Delays the execution of the subsequent remote command by the specified time.
This function is useful, for example to execute an SCPI sequence automatically but
with a defined time delay between some commands.
See Chapter 10.2.4, "How to assign actions to the [★ (User)] key", on page 238.
Setting parameters:
<TimeMs> integer
Wait time in ms
Range: 0 to 10000
*RST: 0
Example:
SYSTem:WAIT 10000
// Waits 10s before resetting the instrument.
*RST
Usage: Setting only
SYSTem subsystem