Remote control commands
R&S
®
SMB100B
570User Manual 1178.3711.02 ─ 08
<RadioTextMsg> Max. 64 characters
Two texts of 64 characters each can be transmitted in a radio
text message.
Note: For group B, the length of a radio text is limited to 32 char-
acters. Special characters in the radio text are entered with a
leading back slash (\) followed by the decimal code of the spe-
cial character according to table E1 of CENELEC.
Example:
STEReo:DIRect "RT=02,0,test text with \217"
Sets the radio text message. 217 denotes the German letter ü.
STEReo:DIRect "SPS=<Time>,<PSN#1>,<PSN#2>,...<PSN#20>"
STEReo:DIRect? "SPS"
Switches program service names (PSN). The program service name automatically
changes after the set time interval.
Parameters:
<PSN> 8 ASCII characters
Max. 20 program service names of eight characters each can be
entered.
Note: Enter the program service names as 8-digit texts. Add
blank spaces, if any.
STEReo:DIRect “SPS=0” stops the transmission of the scroll-
ing PS and starts the transmission of the standard PS.
Setting parameters:
<Time> Time interval in seconds
Range: 00 to 59 s
Example:
STEReo:DIRect "SPS=05,TEST0123,TEST4567"
Transmitts the program service names "TEST0123" and
"TEST4567" alternately, each 5 seconds.
STEReo:DIRect "SRC=<SigSource>"
STEReo:DIRect? "SRC"
(for documentation reasons only)
Selects the signal source.
Use the command [:SOURce]:STEReo:SOURce instead.
Setting parameters:
<SigSource> 0 | 1 | 2 | 3
0
Off
1
External analog (via L and R inputs)
2
External digital
Direct commands for the stereo/rds coder option R&S SMBB-B5